Published by Misasa Medical Center, Okayama University Medical School
Published by Misasa Medical Center, Okayama University Medical School

<Formerly known as>
岡大三朝分院研究報告 (63号-72号) 環境病態研報告 (57号-62号)
岡山大学温泉研究所報告 (5号-56号) 放射能泉研究所報告 (1号-4号)

<Availability>
Some items are not available because of decision by its author or publisher.

温泉治療の血液pH,P(CO(2))並びにP(O(2))に及ぼす影響に関する研究 第l編. 測定機器及び測定条件の吟味

八幡 隆昭 岡山大学温泉研究所温泉医学部門
抄録
Human venous blood pH, P(CO(2)) and P(O(2)) were measured with I. L. Meter. Its reproducibility and response rate were evaluated as well as methods of the procedures. 1) Time required to reach the stability of pH, P(CO(2)), and P(O(2)) reading after the sampie injection was 1~1.5 min., 2~2.5 min. and 45~75 sec., respectively. It is recommended in the measurement of these parameters at the same time that the sample injection starts with P(CO(2)) electrode, followed by pH and P(O(2)) in this order and that readings are recorded in the order of pH, P(O(2)) and P(CO(2)). 2) Range of differences between the two values measured in the interval of 3~5 min. were pH : -0.010~0.020 (mean: 0.003), P(CO(2)) : -1.0~1.0mmHg (mean: 0.4) and P(O(2)) : -1.0~0.0mmHg (mean: -0.5). Their 5% rejection limits were 0.021≧x(o)≧-0.015, 2.0≧x(o)≧-1.2mmHg and 0.3≧x(o)≧-1.3mmHg, respectively. 3) The pH, P(CO(2)) and P(O(2)) of the heparinized venous blood stored in ice water showed no significant changes in 60 min. and they gave practically the same results as the measurement just after shedding.
備考
原著論文 (Original Paper)
ISSN
0369-7142
NCID
AN00032853