JaLCDOI 10.18926/AMO/30998
フルテキストURL fulltext.pdf
著者 Yamada, Teruo| Hirai, Yasuo| Sakano, Shigehiro| Kosaka, Makoto| Tada, Keiichi| Furutani, Sei| Kosaka, Futami|
抄録 <p>The direct determination by gas chromatography of blood levels of anesthetic agents has been difficult because of the water content of blood. In the present study, the method of Yokota et al. (1967) was modified by improving the packing materials of the column, the blood sample vaporizer and the flow-path during analysis. As a result, accurate and reproducible determination of halothane, enflurane and isoflurane dissolved in blood was achieved. With this system, blood in which halothane, enflurane and isoflurane had been dissolved could be analyzed without changing the column between samples. Moreover, each sample was prepared in less than 10 min, and more than 100 consecutive determinations could be made with excellent reproducibility. The coefficient of variation was less than 3.8%.</p>
キーワード determination halothane enflurance isoflurance gas chromatography
Amo Type Article
発行日 1988-08
出版物タイトル Acta Medica Okayama
42巻
4号
出版者 Okayama University Medical School
開始ページ 183
終了ページ 192
ISSN 0386-300X
NCID AA00508441
資料タイプ 学術雑誌論文
言語 English
論文のバージョン publisher
査読 有り
PubMed ID 3177003
Web of Sience KeyUT A1988P884600001